dc.contributor.author | BELOTSERKOVSKII, I. | |
dc.contributor.author | SIBAEV, A. | |
dc.contributor.author | KARAGENOV, D. | |
dc.contributor.author | SIDORENKO, A. | |
dc.date.accessioned | 2024-04-25T06:49:10Z | |
dc.date.available | 2024-04-25T06:49:10Z | |
dc.date.issued | 2024 | |
dc.identifier.citation | BELOTSERKOVSKII, I. et al. Studying the effect of various gases on the output characteristics of thermocouple pressure transducers. In: Quo Vadis – Ethics of the Scientific Research: conf. NANO-2024, event devoted to the 60th anniversary of the Technical University of Moldova, 15-18 April 2024, Chişinău, Republica Moldova: Program and proceedings of the conference, Chişinău 2024, pp. 55-56. ISBN 978-9975-64-422-8. | en_US |
dc.identifier.isbn | 978-9975-64-422-8 | |
dc.identifier.uri | http://repository.utm.md/handle/5014/26901 | |
dc.description.abstract | Experimental studies have shown a real possibility of using thermocouple (and in a broader sense) thermal converters to measure the pressure of various gases in a wide pressure range with a low error, subject to mandatory precalibration of the transducers for each of the gases or their mixture. In addition, a constant pattern of the dependence of the calibration characteristic of the thermocouple transducer for different gases is observed over the entire pressure range of 5 to 5 × 10−3 mmHg. Owing to this feature, a thermocouple transducer can be calibrated for different gases at one pressure point using a strain transducer, the readings of which do not depend on the type of gas; subsequently, certain coefficients can be used for any transducers of this type in a pressure range of 5 to 5 × 10−3 mmHg. | en_US |
dc.language.iso | en | en_US |
dc.publisher | Technical University of Moldova | en_US |
dc.relation.ispartofseries | NANO-2024 "Quo Vadis – Ethics of the Scientific Research", event devoted to the 60th anniversary of the Technical University of Moldova;15-18 April 2024, Chişinău, Republica Moldova | |
dc.rights | Attribution-NonCommercial-NoDerivs 3.0 United States | * |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/3.0/us/ | * |
dc.subject | thermocouple thermal converters | en_US |
dc.subject | thermocouple pressure transducers | en_US |
dc.title | Studying the effect of various gases on the output characteristics of thermocouple pressure transducers | en_US |
dc.type | Article | en_US |